What is the procedure for Programming Version D Circuit Boards
Outlines the procedure for Programming Version D Circuit Boards
Technical Note No. 008
Procedure for Programming Version D Circuit Boards
Date: 1 July 2006
Author: Craig Williford
Requirements
Programming the ozone monitor requires a computer running the "Analog Devices Serial
Downloader" program to download the firmware from the computer to the ozone monitor,
and the rs232 serial cable provided by 2B Technologies. The computer must have a
serial port or a USB port with a USB to Serial converter cable to connect to the rs232
serial cable provided by 2B Technologies.
Procedure
1. Download the Widows Serial Downloader provided by Analog Devices from the
website:
ftp://ftp.analog.com/pub/www/technology/dataConverters/microconverter/wsd_v6_7.exe.
2. Install the Windows Serial Downloader by starting the executable file you have just
downloaded.
3. Start the Windows Serial Downloader program that you just installed.
4. Connect one end of the serial cable provided by 2B Technologies to serial connection
on rear side of ozone monitor and the other end to the computer. Or if the computer does
not have a serial port, connect the other end to the USB to Serial converter cable.
4. Remove the top cover of the ozone monitor. Instructions for removal are located in the
ozone monitor manual.
5. Locate the small buttons labeled “sw1” and “sw2” on the circuit board inside the ozone
monitor.
6. Turn on the ozone monitor.
7. Press and hold the switch labeled “sw2”. While holding “sw2” down, press and
depress the switch labeled “sw1”. Now release “sw2”. This will enter the board into
“Serial Download Mode”.
8. On the computer running the Windows Serial Downloader, locate the “Reset” button
on the program’s window. Press the “Reset” button. If the board is connected correctly
and in “Serial Download Mode”, you will see this message:
“Configuration: COM1, 9600 baud
RESETTING PART: OK”
*If you do not see this message, repeat steps 7 through 8 until you see the correct
message
9. Press the “Download” button located on the Windows Serial Downloader. This will
bring up a window prompting you to choose the file with a .hex extension to be
downloaded. Find the .hex file that was provided by 2B Technologies and select it. A bar
will appear showing the status of the download and the download should only take a few
seconds.
10. The Board is programmed! Turn the ozone monitor off and on again to start the new
program.
11. Enter the menu of the ozone monitor and re-enter all of the parameters (Averaging
time, Calibration parameters, External, Baud Rate, Units). You must do this because the
internal eeprom memory is cleared when the microcontroller is reprogrammed.
Note: If step 8 is unsuccessful, please submit a tech support ticket on our website under customer support for further assistance.