Skip to content
English
  • There are no suggestions because the search field is empty.

How to use the lamp heater control option for Model 202 and 205?

The lamp heater option for the model 202 or 205 can be controlled via serial commands sent from a terminal emulator running on a computer connected to the ozone monitor through a serial cable.

Technical Note No. 026 

Lamp Heater Control for Model 202/205 Ozone Monitor 

Date:  17 August 2010 

Author:  Craig Williford 

Summary:  The lamp heater option for the model 202 or 205 can be controlled 
via serial commands sent from a terminal emulator running on a computer 
connected to the ozone monitor through a serial cable.    

Status:  The status of the heater is output as the last variable of the serial data stream.   
Lamp heater ON = 20 
Lamp heater OFF = 00   

Example:  2,3.1,24.9,900.3,1451,16/10/09,1:45:33,20 

Where the values are separated by commas and the values are defined in this 
header: 
Log#,Ozone,Temperature,Pressure,Flow,Date,Time,Lamp Heater Status 

Commands:  To Control Lamp Heater (lower case): 
l – turn lamp heater ON 
m – turn lamp heater OFF